Product Information
Welcome to the website of Institute of Calibration & Technology Co., Ltd.
■ Application
- Convenient operation and intuitive design with touch display adoption
- High-performance precision measurement available
- Multiple functions can be measured in one piece of equipment
- Can be controlled using serial communication PC program
- Compact design makes it light and easy to move
■ Specification
▶ Voltage measurement
- Voltage range : ± (0.1 ~ 10) kV DC/AC RMS
- Resolution : 0.001 kV
- Accuracy : 0.5 % + 2 digit at DC
0.5 % + 2 digit at AC 50 Hz / 60 Hz
▶ Cutoff current measurement
- Current range : (0.5 ~ 100) mA
- Resolution : 0.001 mA at (0.5 ~ 5) mA range
0.01 mA at (10 ~ 50) mA range
0.1 mA at 100 mA range
- Accuracy : 0.5 % + 1 digit at DC/AC
0.5 % + 2 digit at AC 50 Hz / 60 Hz
▶ Timer time measurement
- Range : (0.3 ~ 9999) s
- Input voltage : 10 V ~ 10 kV
- Resolution : 1 ms
- Accuracy : 0.05 % + 2 ms at DC
0.05 % + 20 ms at AC